Assignment 1.Generally, monochromaters displaying high resolving power are required to contain long focal length optical components. However, an...

Assignment

1.Generally, monochromaters displaying high resolving power are required   to contain long focal length optical components. However, an alternative to this is the incorporation of an Echelle grating which is then operated at higher orders of diffraction.   Unfortunately, at   these   higher orders of diffraction, extensive order overlap is observed.  To overcome this difficulty, a prism is sometimes included as a cross-dispersion element within the monochromater. 

A. For such a cross-dispersing   Echelle monochromater, with   the   following specifications, calculate   the   resolving   power   of   this monochromater at orders 50, 75, 100, and 120 for wavelenths of 200 nm, 400 nm, and 700 nm. (Note: resolution is in each of two dimensions.)

Focal length 0.750 m

Grating width 9.60 cm

Grating height 4.60 cm

Diffraction angle 63.43

Groove density 79 grooves/mm

Prism angle 30

Prism material Fused Silica

Prism base 3.0 cm

Wavelength/nm______

200.0 400.0 700.0

Refractive Index 1.5505 1.4696 1.45515

Dispersion (of the material)

1.08258 1.12698 1.13086

B. Calculate the band pass of the instrument with vertical slit widths of 50, 100, and 200 m and horizontal slits of 10, 25, and 50 m (Hint: they are unrelated).

C. Determine the f/# of the monochromater.

D. Because the internal spherical mirrors of the monochromater are illuminated off-axis, astigmatic images would be predicted to be formed.  With an off-axis angle of 3.0 degrees, compute the distance between the two resulting astigmatic images.  Is that distance significant to prohibit the use of the monochromater (i.e., does the circle of least confusion exhibit tolerable blur)?

E. What would be the dimensions of a multi-point detector to enable the simultaneous measurement of the entire spectrum from 200.0 nm to 650.0 nm? Propose a detector for this purpose. What parameters would need to be changed to use a CCD camera with dimensions of 1.50 cm x 1.50 cm and detecting elements (i.e., pixels) with a separation of 15 μm, center to center?